Model GammaCell 220F Co-60 Self-Contained Irradiator
Features
- Special Form Cobalt 60 Sources
- A Maximum of 888 TBq (24,000) Curies
- Maximum Dose Rate: 33,000 cGy/min
Applications
- Research
Description
The Model GammaCell 220 System is a large capacity (24,000 Curies) Self-Contained Irradiator originally manufactured by AECL or its decendents.
FOSS Therapy Services has re-engineered the GammaCell 220 system and has been awarded a SSDR (Sealed Source & Device Registration) for our Model GammaCell 220F. The GammaCell 220F system is a fully remanufactured irradiator with new electronics, mechanical components, bearings, and NEW Cobalt 60 sources.
Specifications
Sealed Source & Device Registration Number
- CA1418D101S
Special Form Co-60 Sources:
- A maximum of 888 TBq (24,000 Curies)
Maximum Dose Rate:
- 33,000 cGy/min
Sample Chamber
- 6.0″ diameter x 7.9″ high (152 mm x 200 mm)
Access Ports
- Single access port: 1.38″ diameter (35 mm)
Safety Measures
- External Radiation Levels: The GammaCell 220F system has all the same external radiation levels as the standard unit. No changes are made to any shielding or this new device
- Interlocks: The GammaCell 220F system retains all of the original safety interlock as the standard GammaCell 220 unit
Control Panel
- Timer – preset or manual
- ON OFF switches
- Toggle switch to select PRESET or MANUAL (Hours or Minutes) timing
- Keyed power switch with light
- INTERLOCK light (all interlocks are satisfied)
Mechanical & Electrical Specifications:
- Size: 40″ wide x 60″ long x 83.63″ high
(open drawer) (101.6 x 152.4 x 212.4 cm) - Weight: 8300 lbs. (3765 kg)
- Power: 220 V, 50 or 60 Hz, 15 amps
Source Transfer Criteria
- Space Required: 11.5 ft long x 11.5 ft wide x 11.5 ft high (3.5 m x 3.5 m x 3.5m)
- Radiation Secure Area: 13.1 ft square (4 m)
- Foss Therapy has engineered a “portable hot cell” system to perform on-site source replacement and decommissioning. We have reloaded GammaCell 200 systems with > 20,000 Curies. A source reloading project can be completed within one week.
Accessories
- Attenuators: X2, X4, X8
- Cold air/nitrogen input
- Sample chamber
Options
- Can be fitted with Lead/Aluminum insert for Mil-Spec Electronics testing.